Vaisala DMP248 Manual de servicio Pagina 80

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 145
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 79
OPERATING MANUAL_______________________________________________________________
74 _____________________________________________________________________ U263EN-1.4
Temperature
Measurement range -40...+80 °C
for specified Td accuracy 0...+80 °C
Typical accuracy of electronics
at +20 °C (+68 °F) ±0.1 °C
Typical temperature dependence of
electronics 0.005 °C/°C
Temperature sensor Pt 100 IEC 751 1/3 Class B
Relative humidity
Measurement range 0...100 %RH
Accuracy at 20 °C
RH< 10 %RH ± 0.025 %RH + 8.75% of reading
RH> 10 %RH ± 0.7 %RH + 2% of reading
Ppm volume concentration (dry)
Typical measurement range 0...5000 ppm
v
Accuracy at 20 °C, 1013.25 mbar 7.3 ppm
v
+ 8.3% of reading
Outputs
Two analogue outputs selectable 0...20 mA 4...20 mA
0...1 V 0...5 V
0...10 V
Typical accuracy of an analogue
output at +20 °C ±0.05 % full scale
Typical temperature dependence
of an analogue output 0.005 %/°C full scale
Serial outputs RS 232C
RS 485 (option)
current loop (option)