Show User Calibration Coefficients
Use the L command to view the current user calibration coefficients on
the serial line. This command is useful for checking if non-factory
adjustments have been applied for humidity and temperature
measurement.
L<cr>
The output values are as follows:
- RH offset and gain: corrections to RH measurement. These values
change when you calibrate and adjust the humidity measurement.
- T offset and gain: corrections to T measurement. These values change
when you calibrate and adjust the temperature measurement.
For more information on how the adjustments affect these coefficients,
see section Adjustment Points on page 55.
Example (shows default values, no user adjustment done):
l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Set User Calibration Coefficients
Use the LI command to enter values for offset and gain coefficients for
on the serial line. Useful for restoring the factory defaults, or some earlier
state of user adjustment.
CAUTION
Use this command only to restore values you have previously written
down based on the output from the L command, or to restore the default
offset and gain.
LI<cr>
After you enter the command, you will be prompted to enter new values
for the coefficients. The default offset is zero (0), default gain is one (1).
Example (shows user adjusted offset values, overwrites them with 0):
li
RH offset : -4.35084105E-04 ? 0
RH gain : 1.00000000E+00 ?
T offset : -1.62982941E-02 ? 0
T gain : 1.00000000E+00 ?