Chapter 4 ________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 79
Use the UNIT command to change the measurement unit for a quantity.
Specify the quantity and the desired unit for it.
Example:
>unit P mmhg
P : mmHg
P3h : Pa
P1 : Pa
P2 : Pa
DP12 : Pa
HCP : Pa
QFE : Pa
QNH : Pa
>
Use the UNIT ?? command to list the available measurement units for
the quantities.
Measurement Related Commands
TQFE
Use the TQFE command to show or set the temperature for QFE
corrected pressure. The valid range of QFE temperature is -80...+200
°C.
Example:
>tqfe
QFE temp. : 20.00 'C ? 21
>
DPMAX
The serial line command DPMAX shows or sets the maximum
difference pressure allowed between barometer modules. The
command is only available, if more than one barometer module has
been installed.
NOTE
This command changes both the serial output and display units.