Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 33
0604-019
Figure 16 Flange Installation Kit
The following numbers refer to Figure 16 above:
1 = Process or duct wall
2 = Flange
3 = Sealing ring
4 = Probe
NOTE
When the temperature difference between the process or duct and the
surroundings is large, the probe must be installed as deep in the process
or duct as possible. This prevents errors caused by heat conduction in the
probe cable.
DMT346 Probe and Cooling Set Mounting
22.5
ø1
334.5
ø13.5
357
1 2
0604-020
Figure 17 DMT346 Probe Dimensions (in mm)
The following numbers refer to Figure 17 above:
1 =
Stainless steel grid
2 = Cable (length 2 m, 5 m or 10 m)