User's Guide _______________________________________________________________________
112 __________________________________________________________________ M210762EN-G
SMODE
Use the SMODE command to set the user port start-up operating mode.
SMODE [xxxx]<cr>
STOP, RUN, POLL, or MODBUS.
Table 30 Selection of Output Modes
Only with the SEND command.
Only with the SEND [addr]
command.
SEND [addr] and OPEN [addr].
Other commands available after
opening a line to the transmitter
using the OPEN command.
Use with RS-485 buses where
multiple transmitters can share
transmitter using MODBUS
Modbus protocol only; see
Chapter 5, Modbus, on page 137.
Selected output mode will be activated at next reset or power up.
ADDR
Use the ADDR command to set the device address of the transmitter.
Addresses are required for POLL mode and MODBUS mode (serial
Modbus).
ADDR [aa]<cr>
Device address of the transmitter, range 0 ... 255 (default = 0)
Example (changing the transmitter address from 0 to 52):
>addr
Address : 0 ? 52
>