Chapter 5 ______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 45
CHAPTER 5
MAINTENANCE
This chapter provides information that is needed in basic maintenance of
HMP60 and HMP110.
Periodic Maintenance
The measurement accuracy of the HMP60 and HMP110 should be
calibrated yearly. When calibration indicates that accuracy is not within
specification:
- HMP60: change the INTERCAP® Sensor.
- HMP110: adjust the measurement, or have a Vaisala Service Center
adjust it for you.
Light cleaning and filter replacement should be done only when
necessary.
Cleaning
The probe body can be wiped clean with a soft, linf-free cloth moistened
with mild detergent. Do not use solvents or compressed air.
Note that wiping the membrane filter or stainless steel sintered filter may
block its pores and/or deposit residue on the filter. If the filter is heavily
contaminated, replace it.