Selecting Analog Outputs
You can select the output quantities for channels 1 and 2 by turning the
Calibration Enabled/Disabled DIP switch of the transmitter to position
Enabled (up) and the three output selection DIP switches to position
special (all up).
Press button C on the display cover: text "SCAL" appears on the display.
Then press button E until the following display appears:
0505-288
The characters on the second line of the display in this menu correspond to
the quantities according to the table in section Selecting Output Quantities
on page 47.
Select the quantity for Ch 1 with buttons Up and Dn and acknowledge the
selection with button E.
If the transmitter is equipped with two analog channels, select the quantity
for Ch 2 in the same way, for example:
0505-289
Press button C to exit the display command mode or continue by setting
the pressure.
NOTE
Remember to restore the DIP switch settings.