USER'S GUIDE____________________________________________________________________
46 __________________________________________________________________ M211289EN-A
View User Adjustment Parameters
Use the L command to view the current user adjustment parameters that
are affected by the
CRH, RHCAL, CT, and LI commands. This
command is useful for checking the currently applied correction.
L<cr>
The output values are as follows:
- Cp offset and gain: capacitance, adjusted using the
CRH command.
- RH offset and gain: adjusted using the
RHCAL command.
- T offset and gain: adjusted using the
CT command.
Example (shows default values, no user adjustment done):
>l
Cp offset : 0
Cp gain : 1
RH offset : 0
RH gain : 1
T offset : 0
T gain : 1
>
Set User Adjustment Parameters
Use the LI command to set the user adjustment parameters directly. This
command is useful for restoring the desired adjustments, and removing
the effects of user adjustment without having to perform the factory reset.
LI<cr>
After entering the command, you will be prompted to enter new values
one at a time. Press ENTER without giving a value to move to the next
parameter.
Example (shows default values):
>li
Cp offset : 0 ?
Cp gain : 1 ?
RH offset : 0 ?
RH gain : 1 ?
T offset : 0 ?
T gain : 1 ?
>