Technical data
Measured Values
WATER ACTIVITY
Measurement range 0 ... 1 aw
Accuracy (including non-linearity, hysteresis and repeatability)
0 ... 0.9 ± 0.02
0.9 ... 1.0 ± 0.03
Response time
in oil fl ow (typical) <1 min (dry-wet)
MOISTURE
Calculated moisture content in ppm for mineral transformer oil
TEMPERATURE
Accuracy at +20 °C (+68 °F) ± 0.2 °C (0.36 °F)
Operating Environment
Operating temperature -40 ... +60 °C (-40 ... +140 °F)
Oil temperature -40 ... +80 °C (-40 ... +176 °F)
Pressure range
metal version up to 200 bar
plastic version up to 40 bar
Oil fl ow some ow recommended
Options and accessories
Stainless steel fi lter (standard) 221494SP
Stainless steel fi lter for high fl ow (>1 m/s) 221493SP
Connection cable for MM70 hand-held meter 219980
USB serial interface cable 219690
Sealing ring set (U-seal) ISO G1/2, 3 pcs 221525SP
Sealing ring set (copper) ISO G1/2, 3 pcs 221524SP
ISO 1/2’ plug 218773
NPT 1/2’ plug 222507
Sampling cell DMT242SC
Sampling cell w. Swagelok connectors DMT242SC2
Connection cable
2 m (6.5 ft), M8 snap-on 211598
0.32 m (1 ft) Shielded, M8 threaded HMP50Z032
3.0 m (9.8 ft), Shielded, M8 threaded HMP50Z300
5.0 m (16.4 ft), Shielded, M8 threaded HMP50Z500
10 m (32.8 ft), Shielded, M8 threaded HMP50Z1000
3 m, Shielded, connector 90° angle 221739
5 m, Shielded, connector 90° angle 221740
M8 threaded, Ch1 signal + Ch2 LED MP300LEDCBL
Outputs
Analog outputs (two channels)
current output 0 ... 20 mA, 4 ... 20 mA
voltage output 0 ... 5 V, 0 ... 10 V
Alarm level indication by analog signal user selectable
Digital outputs RS-485
Ref. B210755EN-B ©Vaisala 2009
This material is subject to copyright protection, with all
copyrights retained by Vaisala and its individual partners. All
rights reserved. Any logos and/or product names are trademarks
of Vaisala or its individual partners. The reproduction, transfer,
distribution or storage of information contained in this brochure
in any form without the prior written consent of Vaisala is strictly
prohibited. All specifi cations — technical included — are subject
to change without notice.
For more information, visit
www.vaisala.com or contact
us at sales@vaisala.com
General
Sensor HUMICAP
®
Cable connections (2 ports) M8, 4 pin
Minimum operating voltage with
RS-485 output 14 ... 28 VDC
voltage output 16 ... 28 VDC
current output 22 ... 28 VDC
Supply current
normal measurement 20 mA + load current
External load for
voltage output min. 10 kOhm
current output max. 500 Ohm
Housing material
metal AISI 316L
plastic PPS + 40% GF
Mechanical connections with bonded seal ring (washer)
metal version G 1/2’ ISO or NPT 1/2’
plastic version G 1/2’ ISO
Housing classifi cation IP65 (NEMA 4)
Storage temperature range -40 ... +80 °C (-40 ... +176 °F)
Weight
with plastic housing 65 g (2.3 oz)
with metal housing 200 g (7 oz)
Complies with EMC standard EN61326-1, Electrical equipment
for measurement control and laboratory use - EMC requirements;
Industrial environment