Chapter 3 ________________________________________________________________ Installation
VAISALA ________________________________________________________________________ 23
Ball Valve Installation Kit for MMT318
The optional ball valve installation kit (Vaisala order code:
BALLVALVE-1) is preferred when connecting the probe to a
pressurized process or pipeline. Use the ball valve set or a 1/2" ball valve
assembly with a ball hole of Ø14 mm or more.
If you install the sensor head (Ø 12 mm) in a process pipe, please note
that the nominal size of the pipe must be at least 1 inch (2.54 cm). Use a
manual press handle to press the sensor head into the pressurized (< 10
bar) process or pipeline.
0507-043
Figure 8 MMT318 Ball Valve Assembly Installation
The following numbers refer to Figure 8 above:
1
=
Manual press tool
2
=
Handle of the ball valve
3
=
Probe
4
=
Process chamber or pipeline
5
=
Groove on the probe indicates the upper adjustment limit
6
=
Filter
7
=
Ball of the ball valve
8
=
Fitting screw
Vista de pagina 22