CHAPTER 6_______________________________________________ CALIBRATION AND ADJUSTMENT
VAISALA _________________________________________________________________________ 35
MPC Multipoint pressure corrections
MPC x <cr>
where:
x = ON or OFF
The MPC command is used to activate or deactivate the multipoint
corrections. Plain command MPC outputs the corrections in use.
The multipoint corrections are protected with switch SW4 (see FIGURE
6-1). Turn the switch SW4 to write ENABLE position (ON) to be able
to turn the multipoint corrections on or off.
NOTE
Changing the multipoint corrections on or off will automatically
cancel the previous date of calibration of the transmitter (see CALD
command).
Example of multipoint adjustments performed on a transmitter with
one pressure transducer:
>mpc off <cr>
Multipoint adj: OFF
>mpc on <cr>
Multipoint adjustments ON
>mpc<cr>
Reading Correction Reading Correction
499.660 -0.110*******.**********.***
599.110 -0.080*******.**********.***
698.580 -0.060*******.**********.***
800.950 -0.030*******.**********.***
900.400 0.010*******.**********.***
947.200 0.020*******.**********.***
999.840 0.050*******.**********.***
1099.090 0.070*******.**********.***
Multipoint adj: ON
>
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).