User's Guide ______________________________________________________________________
134 _________________________________________________________________ M210796EN-A
adjustment information (date and text) to the memory of the
transmitter, see the commands CTEXT and CDATE.
13. Press the ADJ button on the motherboard to close the
adjustment mode.
14. Take the probe out of the reference conditions and replace the
filter.
Relative Humidity Adjustment After Sensor
Change
Using Display/Keypad
When using the optional display/keypad, follow the instructions on
Using Display/Keypad on page 131 on page but select Adj. for new
RH sensor (instead of 1-point/ 2-point adjustment).
Using Serial Line
After sensor change, carry out the procedure as described in previous
sections. Just replace the CRH command with the FCRH command.
FCRH
Example:
>FCRH
RH : 1.82 1. ref ? 0
Press any key when ready...
RH : 74.22 2. ref ? 75
OK
>
The OK indicates that the calibration has succeeded.