User's Guide ______________________________________________________________________
24 __________________________________________________________________ M210681EN-B
The Data Logging module must be running continuosly to enable
updating of data from the data collection system to the data storage on
the hard disk.
The Data Logging module stores data only if the data is changed. By
default, there is no minimum limit for data change, but it is possible to
define a minimum change with the Deadband parameter. All data in
each log definition is stored if any of the items change.
The Data Logging module also supports FTP and E-mail data transfer.
However, these features are optional and require installation-specific
settings in the YourVIEW initialization files before they can be used.
Changing the Configuration
The settings for the Data Logging module are saved in a file called
logger.cfg.
NOTE
The old file is overwritten when changes are saved. Therefore, make
a backup copy of the original file.
Changes in the configuration of YourVIEW Data Logging must be
made while the Receiver is running. However, the changes will not be
effective until you restart the module.