Vaisala Drycap DMT152 Guía de usuario Pagina 55

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 70
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 54
Chapter 5 ________________________________________________________ Serial Line Operation
VAISALA ________________________________________________________________________ 53
Other Commands
Opening the Transmitter in POLL Mode
Use the OPEN command to connect to a transmitter that is in POLL
mode.
OPEN [aa]<cr>
where
aa
=
address (0 ... 99)
Example:
open 1
DMT152 1 line opened for operator commands
Closing the Connection to a Transmitter
in POLL Mode
The CLOSE command closes the connection to a transmitter.
CLOSE<cr>
Example:
close
line closed
Display Command List
HELP<cr>
Set Time
Use the TIME command to view and set the transmitter clock. The clock
does not have a battery backup – it is reset to 00:00:00 when the
transmitter is powered up or reset.
TIME [hh mm ss]<cr>
where
hh
=
Hours (24-hour clock)
mm
=
Minutes
ss
=
Seconds
Vista de pagina 54
1 2 ... 50 51 52 53 54 55 56 57 58 59 60 ... 69 70

Comentarios a estos manuales

Sin comentarios