Application note
Vaisala WXT520 and Nokeval 7470 V3.0-
Juha Hämäläinen / Nokeval Oy 16.10.2013
Vista de pagina 0
1 2 3 4 5 6 ... 13 14

Indice de contenidos

Pagina 1 - Application note

Application noteVaisala WXT520 and Nokeval 7470 V3.0-Juha Hämäläinen / Nokeval Oy 16.10.2013

Pagina 2 - How to go

CONNECTING 7470 FOR RS-232/485 CONFIGURATIONUsing RS-232 or RS-485 for configuring the 7470 is necessary only if a programming cable is not available.

Pagina 3

CONFIGURING THE 7470ConnectingOpen a MekuWin connection to 7470, using the serial number or the serial port of the programming cable, SCL protocol, 96

Pagina 4 - CONFIGURATION TOOL

when it is 50, set Range=4-20mA, Lo=0, and Hi=50. The 7470 does not know the measurement unit – it is defined in the Vaisala configuration.Finally cli

Pagina 5

WIRINGRS-485Separate serial and power wiringCombined wiringSince Vaisala does not have an isolation between serial and power ports, one wire can serve

Pagina 6

Wiring with the Vaisala M12 cableWXT5201245674701112+-PowerM12GreyBlueRed (or clear)Brown3

Pagina 7 - SOFTWARE

GeneralNokeval 7470 analog output unit can be programmed to interpret messages from Vaisala WMT and WXT weather sensors. This requires configuring bot

Pagina 8

Vaisalaservice cableavailableNo YesPOL cableavailableNo YesWant to useSicalaNo YesConfigure Vaisalaw ith terminalsoftw areConfigure Vaisalaw ith Si

Pagina 9

CONFIGURING VAISALA WITH VAISALA CONFIGURATION TOOLConnectingLaunch the Vaisala configuration tool. Select File/Connection Setup. Select your serial p

Pagina 10 - CONFIGURATION

Message settingsSelect the parameters required. The Composite message section is the only that matters. Do not select more than 11 parameters, or the

Pagina 11 - CONFIGURING THE 7470

Profile managerVaisala Configuration Tool has a profile manager that allows viewing all the settings at once and saving them to a file. The profile ma

Pagina 12 - Troubleshooting

CONFIGURING VAISALA WITH A TERMINAL SOFTWARESetting up the terminal softwareOpen HyperTerminal (or any other terminal software) and create a session t

Pagina 13 - Combined wiring

And switch the station off for 30 s. Change the HyperTerminal to the new baud rate (Call/Disconnect, File/Properties/Configure and Call/Call again).Co

Pagina 14

CONFIGURING VAISALA WITH SICALAThe 7470 allows sending commands to the RS-232/485 bus from the POL (programming) port. Nokeval has planned to make a P