Chapter 5 _________________________________________________________________ Operation
VAISALA ________________________________________________________________________ 51
Select Unit
Use the UNIT command to select metric or non-metric output units.
UNIT [x]<cr>
where
x
=
Selects the unit type to output:
m = metric units, for example, Celsius
n = nonmetric units, for example, Fahrenheit
Example:
unit m
Units : Metric
View User Adjustment Parameters
Use the L command to view the current user adjustment parameters that
are affected by the LI command and the MI70 (Tdf offset only). This
command is useful for checking the currently applied correction.
L<cr>
The output values are as follows:
- RH offset and gain
- T offset and gain
- Tdf offset and gain
Example (shows default values, no user adjustment done):
l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: 0.00000000E+00
Tdf gain : 1.00000000E+00
Example (shows that a -1.2 °C Tdf offset has been applied by the user):
l
RH offset : 0.00000000E+00
RH gain : 1.00000000E+00
T offset : 0.00000000E+00
T gain : 1.00000000E+00
Tdf offset: -1.20000000E+00
Tdf gain : 1.00000000E+00