USER'S GUIDE ____________________________________________________________________
54 ___________________________________________________________________ M211435EN-E
SERI [b p d s]<cr>
where
b
=
baud rate (4800, 9600, or 19200)
p
=
parity (n = none, e = even, o = odd)
d
=
data bits (7 or 8)
s
=
stop bits (1 or 2)
Example (set serial line to 9600 baud, no parity, 7 data bits, and 1 stop
bit):
seri 9600 e 7 1
SCI Baud P D S: 9600 E 7 1
Set Transmitter Address
To operate in the POLL mode, the transmitter must have an address. If
multiple transmitters share the same serial line, each transmitter must
have a different address.
For a description of the serial interface modes, see section Set Serial Line
Operating Mode on page 53.
ADDR [address]<cr>
where
address
=
Transmitter address, range 0 ... 255.
Example:
addr
Address : 0 ?
Vista de pagina 55