Vaisala HMP243 Guía de usuario Pagina 117

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 130
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 116
HMP240
M210300en Appendix 5: Calculation formulas
115
(TP5,TP6,TP7 and TP8 for the humidity probe and TP1, TP3 and TP4 for the
additional temperature probe). If you cannot find any faults, send the
transmitter to Vaisala for repair.
E53 U1 y-value out of range
- the humidity measurement cannot be completed. Check the HUMICAP
sensor and humidity measurement wires and their solderings on the
component board (TP9,TP10 and TP11/15). If you cannot find any faults, send
the transmitter to Vaisala for repair.
E54 U2 y-value out of range
- the measurement hybrid is probably damaged. Send the transmitter to
Vaisala for repair.
The LEDs display these error types as follows:
¤¡¡¡
RH channel y-value out of range
¡¤¡¡
T channel y-value out of range
Vista de pagina 116
1 2 ... 112 113 114 115 116 117 118 119 120 121 122 ... 129 130

Comentarios a estos manuales

Sin comentarios