Vaisala HMP243 Guía de usuario Pagina 42

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 130
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 41
HMP240
User's Guide M210300en
40
5.4.3 Operating the transmitter via the serial bus
5.4.3.1 Setting the serial interface
SMODE xxxx<cr>
xxxx = STOP, RUN or POLL
In STOP mode: measurements output only by command, all commands can be
used
In RUN mode: outputting automatically, only command S can be used
In POLL mode: measurements output only with command SEND. When in
POLL mode, the output mode is changed as follows:
OPEN aa <cr>
SMODE xxxx<cr>
aa = address of the transmitter
xxxx = STOP, RUN or POLL
The OPEN command sets the bus temporarily in STOP mode so that the
SMODE command can be given.
Example:
>SMODE STOP <cr> setting STOP mode
Serial mode : STOP
Vista de pagina 41
1 2 ... 37 38 39 40 41 42 43 44 45 46 47 ... 129 130

Comentarios a estos manuales

Sin comentarios