HMP140A SERIES
Operating Manual U175en-1.5
12 1999-12-08
6. TECHNICAL DATA
6.1 Relative humidity
Measurement range 0...100 %RH
Accuracy (including non-linearity and repeatibility)
when calibrated against salt solutions (ASTM E104-85):
0...90 %RH +2 %RH
90...100 %RH +3 %RH
Temperature dependence:
2
1.5
1
0.5
0
-0.5
-1
-1.5
-2
dRH (%RH)
-20
-10
0102030
40
50
6
0
T (°C)
Response time (90%) 15 s at +20 °C in still air (with sintered filter)
Maximum error between
the display and the outputs 0.2 %RH
Humidity sensor HUMICAP
180
6.2 Temperature
Measurement range -20...+60 °C
Typical accuracy at:
+20 °C +0.2 °C
-10...+40 °C +0.3 °C
-20...+60 °C +0.4 °C
Temperature sensor Pt 1000 IEC 751 Class B