User's Guide ______________________________________________________________________
168 _________________________________________________________________ M210566EN-H
Example of Climate Chamber Installation
0507-016
Figure 93 Climate Chamber Installation (not Available from
Vaisala)
The following numbers refer to Figure 93:
1 = PTFE sleeve
2 = Cable gland, for example: AGRO 1100.12.91.065
3 = Stainless steel cable tie or similar fastener
l
4 = To be sealed (silicone)
5 = Temperature probe
6 = Relative humidity probe
7 = HMP247CG, Cable gland AGRO (available from Vaisala)
NOTE
Let the cables hang loosely to prevent condensed water running to the
probe.