Vaisala HUMICAP HMT330 SERIES Guía de usuario Pagina 40

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 178
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 39
User's Guide ______________________________________________________________________
40 __________________________________________________________________ M210566EN-H
Probe Mounting
In humidity measurement and especially in calibration it is essential that
temperature of the probe and measuring environment is the same. Even a
small difference in temperature between the environment and the probe
causes an error. As the curve below shows, if the temperature is +20 °C
and the relative humidity 100 %RH, a difference of ±1 °C between the
environment and the probe causes an error of ±6 %RH.
The graph below illustrates the measurement error at 100 %RH when the
difference between the ambient and sensor temperature is 1 °C.
0507-023
Figure 22 Measurement Error at 100 %RH
Vista de pagina 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 177 178

Comentarios a estos manuales

Sin comentarios