USER'S GUIDE____________________________________________________________________
28 __________________________________________________________________ M210474EN-B
SERI Serial line settings
SERI b p d s ↵
b = bauds (300, 600, 1200, 2400, 4800, 9600,19200)
p = parity (n = none, e = even, o = odd)
d = data bits (7 or 8)
s = stop bits (1 or 2)
The settings can be changed one parameter at a time or all parameters
at once:
>SERI O ↵ changing parity only
4800 O 7 1
>SERI 600 N 8 1 ↵ changing all parameters
600 N 8 1
ADDR Setting the transmitter address for use in POLL-mode.
ADDR aa ↵
aa = address (0...99)
Example:
>addr
Address : 0
>addr 1
Address : 1
OPEN Temporarily opens transmitter from POLL-mode to
receive serial commands
OPEN nn ↵
nn = address of the transmitter (0...99)
The OPEN command sets the bus temporarily in STOP mode so that
the SMODE command can be given.
Example:
open 4
Device: 4 line opened for operator commands
>