USER'S GUIDE____________________________________________________________________
58 __________________________________________________________________ M210474EN-B
PPM- calculation setting
OIL Oil calculation setting
(Calculation Model with Average
Coefficients
Give a command OIL ON when you are measuring moisture in oil and
you want to have ppm-output.
OIL x ↵
x= ON/OFF
Example:
>oil on
Oil ppm : ON
>oil
Oil ppm : ON
Oil[0] : -1.66269994E+03
Oil[1] : 7.36999989E+00
>
OIL Changing the calculation coefficients (Calculation
Model with Oil Specific Coefficients)
1. Press the blue adjustment button (see picture on page 26) to enable
feeding of coefficients.
2. Give a command OIL.
OIL ↵
3. Type the first coefficient after the question mark and press
ENTER.
4. Type the second coefficient after the second question mark and
press
ENTER.
5. Reset the transmitter by giving the command RESET Turn off and
on the power to return the transmitter to the normal mode.
Example:
>oil
Oil ppm : ON
Oil[0] : -1.66269994E+03 ?
Oil[1] : 7.36999989E+00 ?
>
Comentarios a estos manuales