grees celsius which was kept within a period of thirty minutes before skipping to the
next temperature level.
Figure 5.1: Temperature Measurement as a Function of Time
The blue and green line represent the two sensors, whereas the red line expresses the
reference device. It is striking that there is almost no difference in readings among
the sensors. However, there are varieties in response time in comparison to the ref-
erence device. The reason for that is the use of the ZE21-type sintered high-grade
steel filter which leads to a change in dynamic behavior because of its fine pored
state to protect the sensing element for salt mist.