HMI41 AND PROBES
Operating Manual M210316EN-A
56
3.2 Setting the measurement duration for automatic data collecting
RH
set min max hyst hold Lo bat Hi
Pabs Pa
g/m
T
Td °C°F
°C°F
%
Turn the indicator on. Within 1 - 2 seconds press
the button HOLD until the text REC AUTO appears,
then release the button.
Press ENTER when the text REC AUTO is displayed.
max
Previously set duration time appears. If the previ-
ously set duration is too long for the current battery
charge, the calculated longest duration time possible
for the current battery charge appears instead; this is
also indicated with the text MAX.
bat
Set the duration time with buttons ▲ or ▼. The
duration time can be set from 15 minutes to 7 days.
Text BAT on the display indicates that the battery
charge is not sufficient for the duration time
selected; select a shorter duration time.
Turn the indicator off or press ENTER for setting
the measurement interval.
3.3 Setting the measurement interval for automatic data collecting
min
Previously set measurement interval appears. If
the previously set interval is too long for the
indicator's current memory capacity, the calcu-
lated shortest possible interval appears instead;
this is also indicated with the text MIN.
Select the interval with buttons ▲ or ▼.
Text LO on the display indicates that there is not
enough memory left for the chosen interval;
select a longer interval.
Press ENTER to activate the measurements in
the automatic data collecting mode.
RH
set
T°C
%
Press ON/OFF if you wish to stop the automatic
data collecting.