HMI41 AND PROBES
M210316EN-A Operating Manual
57
3.4 Reading the measurement results
RH
set min max hyst hold Lo bat Hi
Pabs Pa
g/m
T
Td °C°F
°C°F
%
Turn the indicator on. Within 1 - 2 seconds press
the button HOLD until the text REC AUTO appears,
then release the button.
Press twice the button MODE.
Press ENTER.
RH
T°C
%
Numbers on the first line indicate the stored reading
of the quantity shown; numbers on the left of the
second line indicate the sequence number of the
stored reading in the indicator memory. Numbers on
the right of the second line indicate the temperature
measured simultaneously with the first line reading.
The decimals of the T reading can be seen by
pressing the button ENTER:
RH
T°C
%
The display returns to the previous display after a
couple of seconds.
T
T
d
°C
°C
You can change the quantity on the first line by
pressing MODE.
RH
T°C
%
To scroll all measurement results, press ENTER.
An arrow is displayed on the higher righthand
corner. Press HOLD while the arrow is displayed.
Note that the sequence number changes.