User's Guide _______________________________________________________________________
110 __________________________________________________________________ M210796EN-G
Using Serial Line
PRES and XPRES
Command XPRES must be used if the value is changed frequently (for
example, by an automatic system that updates the value). Its value is not
retained at reset, and when set to 0, last value set with PRES is used
instead. Use the serial line and do the following:
PRES [aaaa.a]<cr>
XPRES [aaaa.a]<cr>
Absolute process pressure (hPa)
Example:
>pres
Pressure : 1013.00 hPa ?
>pres 1010
Pressure : 1010.00 hPa
>
Table 31 Conversion Factors for Pressure Units
Example:
29.9213 inHg = 29.9213 x 33.86388 = 1013.25 hPa
PFIX
Use the PFIX command to select either P inputted or P measured.
- When PFIX is On, fixed PRES value is used
- When PFIX is Off, measured PRES value is used