Vaisala BAROCAP PTB330 Guía de usuario Pagina 119

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 144
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 118
Chapter 6 ___________________________________________________Calibration and Adjustment
VAISALA ______________________________________________________________________ 117
5. For linear adjustment, enter the command LCP1 (LCP2,
LCP3), and enter the new adjustments when prompted. See the
first example.
For 1-point offset adjustment, enter the command
LCP1 (LCP2,
LCP3), and enter the measured and reference values in the first
slots, and then enter the same values increased by one in the
second slots. See the second example.
6. Enter the command
LCP1 ON (and LCP2 ON, LCP3 ON, if
applicable) to activate the new adjustments.
LCP1 [x/y] [z]
where:
x = ON
y = OFF
z = ?
Examples:
Linear adjustment:
>lcp1
1. Reading ? 981.2
1. Reference ? 980.0
2. Reading ? 1102.1
2. Reference ? 1100.0
>
Offset adjustment:
>lcp1
1. Reading ? 981.2
1. Reference ? 980.0
2. Reading ? 982.2
1. Reference ? 981.0
>
NOTE
The new linear/offset adjustments cancel the previous linear/offset
adjustments as well as the valid date of calibration of the transmitter.
A new calibration date and note can be set with the CDATE and
CTEXT commands.
Vista de pagina 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 143 144

Comentarios a estos manuales

Sin comentarios