Chapter 4 ________________________________________________________________ Operation
VAISALA _______________________________________________________________________ 85
The settings can be changed one parameter at a time or all parameters
at once:
>seri
Baud P D S : 9600 N 8 1
>seri o
Baud P D S : 9600 O 8 1
>
SMODE
Use the command SMODE to set the user port start-up operating
mode.
SMODE [x]
where:
x = STOP, RUN, POLL, SEND, or PA11A
Table 16 Selection of Output Modes
Value Description
STOP Use STOP protocol on the User Port by default.
RUN Use RUN protocol on the User Port by default.
POLL Use POLL protocol on the User Port by default.
SEND Use SEND protocol on the User Port by default.
PA11A Use STOP protocol with fixed PA11A output message
on the User Port by default (see PA11A Emulation
Mode on page 139).
Selected output mode is activated after power outages.