Vaisala FD12P Guía de usuario Pagina 62

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 154
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 61
User's Guide _______________________________________________________________________
62
______________________________________________________ M210296en-A
The FD12P answers
LINE OPENED FOR OPERATOR COMMANDS
If no input is given within 60 seconds, the FD12P closes the line
automatically.
CLOSE Command
Line can be released to automatic data messages by typing
CLOSE
The FD12P answers
LINE CLOSED
Automatic Message Sending
In the automatic (CLOSEd) mode, the FD12P sends the predefined
message at selected intervals. The automatic message is selected by
the AMES command.
AMES Message_number Message_interval
where
Message_number
- The valid range is 0 ... 7, refers to section Message Types on
page 63.
- Selects the corresponding message. Any negative message
number is converted to 0.
- If only the message number is given, the previous message
interval setting is used.
Vista de pagina 61