User's Guide _______________________________________________________________________
22 ___________________________________________________________________ M211060EN-F
Mounting the HMP60, HMP110, HMP110D and
HMP110T Probes
HMP60, HMP110, HMP110D and HMP110T are designed to be
mounted from the M12 thread on the probe body or from the smooth part
of the probe body. For a convenient installation, use the optional
installation accessories:
- Use the plastic mounting nuts to hold the probe in a through-wall
installation.
- Use the probe mounting clamp to hold the probe on a wall.
- Use the probe mounting flange to hold the probe in a through-wall
installation.
NOTE
Avoid placing the probe in a place where condensation can run onto the
sensor.
Probe Assembly with Duct Installation Kit
0505-177
Figure 18 Assembly of the Probe with Duct Installation Kit
The following numbers refer to Figure 18 above:
HMP60, HMP110, HMP110D or HMP110T probe
1. Slide the probe cable through the duct installation kit plastic pipe.
2. Attach the cable to the probe.
3. Attach probe assembly to the duct.