User's Guide _______________________________________________________________________
60 ___________________________________________________________________ M211060EN-F
10. Confirm the adjustment, press YES (by pressing NO you return to
adjustment mode display and no changes are made).
11. Adjustment has been carried out. Press BACK to exit the
adjustment mode, and press EXIT to return to the basic display.
One-Point Adjustment of Temperature
Measurement
Temperature adjustment can be done if there is reason to believe that the
adjustment is changed. In a 1-point adjustment, make sure the reference
condition represents the measuring environment.
To perform a one-point adjustment to the temperature measurement of
the HMP110, HMP110D or HMP113 using the MI70 indicator, you
need:
- MI70 indicator.
- Connection cable for MI70 indicator (219980).
- One known and stable temperature reference.
1. Connect the probe to Port I of the MI70 indicator.
2. Turn on the MI70 indicator.
3. Start the adjustment sequence from Main menu > Functions >
Adjustments.
4. MI70 notifies you that automatic power off is disabled during
adjustment mode, press OK to acknowledge.
5. Select the T parameter when prompted.
6. Now the adjustment mode is on, press ADJUST to select the
adjustment method.
7. Select 1-point adjustment, press SELECT.
8. Set the probe to a reference temperature. You can follow the
stabilization from the GRAPH display. Press READY when the
reading is stabilized in the reference.
9. Give the reference temperature value by using the arrow buttons.
Press OK.
10. Confirm the adjustment, press YES (by pressing NO you return to
adjustment mode display and no changes are made).
11. Calibration is carried out. Press BACK to exit the adjustment mode
and EXIT to return to the basic display.