Vaisala HUMICAP HMP41 Guía de usuario Pagina 27

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 96
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 26
Chapter 3 _______________________________________________________________ Installation
VAISALA _______________________________________________________________________ 25
1. Open the transmitter cover by pressing the releasing button at the
bottom side of the transmitter.
2. Insert the signal wires through the selected cable gland/conduit
fitting in the bottom of the transmitter or alternatively through the
rubber grommet at the back side of the transmitter.
3. Connect the wires as indicated in Figure 11, Figure 12, and in
Table 3 below. Suitable wire size is between 0.5mm
2
and 1.5mm
2
.
4. Switch on the RS-485 bus termination switch if necessary. For
more information on the RS-485 bus termination switch see section
5. Close the cover by keeping it slightly tilted and first attaching it to
the fixing snaps at the top of the enclosure base and then pushing
the lower part of the cover firmly forward until it locks. The
transmitter is ready for use.
S
G
N
D
C
H
1
C
H
2
S
H
I
E
L
D
V
I
N
-
V
I
N
+
V
V
10 ... 35 VDC
24 VAC
+
-
1011-163
Figure 11 Power Supply And Voltage Outputs
NOTE
15 ... 35 VDC or 24 VAC is needed when using the 0 ... 10 V output.
Vista de pagina 26