Other Commands
Set/Show Displayed Quantities
After you have defined the quantities with the CALCS command, use the
DSEL command to show or set the displayed quantities. Number of
parameters determine the display mode (ie. one or two rows). If one
quantity is given, display will switch to one row mode. If two quantities
are given, two row mode is used. Valid quantities are: RH, T, TD, TDF,
A, X, H, TW, PWS and PW.
DSEL [q1 q2]<cr>
where
q1 q2
= Quantities to be shown on transmitter display
Syntax for single-parameter transmitters:
DSEL [q]<cr>
where
q
= Quantity to be shown on transmitter display
Examples:
>dsel
1. quantity : RH
2. quantity : T
>
>dsel rh t
1. quantity : RH
2. quantity : T
>
>dsel t
1. quantity : T
>