User's Guide ______________________________________________________________________
32 __________________________________________________________________ M211280EN-A
NOTE
Single-parameter transmitters HMT130H and HMT130T have only one
quantity (T for HMT130T, and RH, TD, TDF, A, X, H, TW, PWS, PW
for HMT130H.)
Set/Show Calibration Date
Use the CDATE command to show or set the calibration date field.
(Calibration refers to the calibration of the HMT130's analog outputs.)
CDATE [date]<cr>
where
date
= Calibration date in format yyyy-mm-dd
Examples:
>cdate
Calibrated : "NOT SET"
>
>cdate 2009-06-22
Calibrated : "2009-06-22"
>
Set/Show Calibration Information
Use the CTEXT command to show or set the calibration info field.
(Calibration refers to the calibration of the HMT130's analog outputs.)
CTEXT [location]<cr>
where
location
= Freely defined text string typically indicating the
calibration location
Examples:
>ctext
Calib. info: "VAISALA HELSINKI"
>
>ctext "Calibration lab 2"
Calib. info: "Calibration lab 2"
>