Vaisala PTB220 Series Guía de usuario Pagina 56

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 113
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 55
USER'S GUIDE_______________________________________________________________________
50 ___________________________________________________________________ M210194EN-A
Example of closing a barometer:
>close <cr>
line closed
SEND mode
In the SEND mode the barometer will automatically output one
message at power-up or when triggering the barometer on using pin 4
of the barometer. The trigger signal must be a TTL level signal. TTL
HIGH (5 VDC) turns the barometer ON and TTL LOW (0 VDC) turns
the barometer OFF.
Examples of outputting with the prompt (ECHO ON) and without the
prompt (ECHO OFF):
>echo <cr>
Echo : ON
>smode send <cr>
Serial mode : SEND
>reset <cr>
1013.25 hPa
>
>echo off <cr>
Echo : OFF
reset <cr> (text invisible)
1013.25 hPa
Example of outputting a minimum amount of (six) characters by
excluding the <cr> and <lf> characters, the pressure unit and the
decimal point with the FORM command:
>form <cr>
4.2P""UUUU #r #n
? 6.0 P <cr>
>unit Pa
P unit : Pa
>smode send <cr>
Serial mode : SEND
>echo off <cr>
Echo : OFF
reset <cr> (text invisible)
101325
or
(power-up)
101325
Vista de pagina 55