In case of one transducer, the listing looks as follows:
>corr <cr>
Linear adjustments OFF
Multipoint adjustments ON
Reading Correction
499.720 -0.070
599.200 -0.080
698.710 -0.010
801.120 -0.010
900.610 -0.030
947.420 -0.020
1000.100 -0.040
1099.580 -0.040
Calibration date 1997-01-01
>
In case of two transducers, the listing looks as follows:
>corr <cr>
Linear adjustments OFF
Multipoint adjustments ON
Reading Correction Reading Correction
499.720 -0.070 499.710 -0.060
599.200 -0.080 599.170 -0.050
698.710 -0.010 698.670 -0.070
801.120 -0.010 801.090 -0.080
900.610 -0.030 900.570 -0.090
947.420 -0.020 947.380 -0.080
1000.100 -0.040 1000.050 -0.090
1099.580 -0.040 1099.500 -0.060
Calibration date 1997-01-01
>
In case of three transducers, the listing looks as follows:
>corr <cr>
Linear adjustments OFF
Multipoint adjustments ON
Reading Correction Reading Correction Reading Correction
499.720 -0.070 499.710 -0.060 499.640 -0.090
599.200 -0.080 599.170 -0.050 599.160 -0.040
698.710 -0.010 698.670 -0.070 698.640 -0.040
801.120 -0.010 801.090 -0.080 801.070 -0.060
900.610 -0.030 900.570 -0.090 900.560 -0.080
947.420 -0.020 947.380 -0.080 947.390 -0.090
1000.100 -0.040 1000.050 -0.090 1000.060 -0.100
1099.580 -0.040 1099.500 -0.060 1099.540 -0.100
Calibration date 1997-01-01
>