Vaisala PTB220 Series Guía de usuario Pagina 70

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 113
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 69
USER'S GUIDE_______________________________________________________________________
64 ___________________________________________________________________ M210194EN-A
CALD Storing the new date of calibration
CALD yyyy-mm-dd <cr>
This command is used to store the new valid date of calibration in the
memory of the barometer.
Example:
>cald <cr>
Calibration date ????-??-??
>cald 1997-01-01 <cr>
Calibration date 1997-01-01
>
Changing the status (ON/OFF) of the linear or multipoint corrections
or entering new linear or multipoint corrections will automatically
cancel the valid date of calibration.
NOTE
Always remember to return the switch SW4 to write DISABLE
position (OFF).
Vista de pagina 69
1 2 ... 65 66 67 68 69 70 71 72 73 74 75 ... 112 113

Comentarios a estos manuales

Sin comentarios