USER'S GUIDE____________________________________________________________________
108 _________________________________________________________________ M210796EN-E
ERRS
Use the command ERRS to display transmitter error messages, see Error
States on page 137 and Table 33 on page 138.
Example:
>errs
No errors
>
Example:
>ERRS
FAIL
Error: Temperature measurement malfunction
Error: Humidity sensor open circuit
>
VERS
Use the command VERS to display software version information.
Example:
>vers
PTU300 / 3.01
>
Resetting Transmitter By Using Serial
Line
RESET
Resets the device. The user port switches to start-up output mode
selected with command SMODE.
Locking Menu/Keypad by Using Serial
Line
LOCK
Use the LOCK command to prevent the user from entering the menu
using the keypad, or to lock the keypad completely. You can optionally
set a 4-digit PIN code, for example 4444.