Configuring Loggers
Vaisala Veriteq vLog 4.5 33
Setting Logger Sample Timing
There are four main steps in setting the Logger Sample
Timing:
a Set the Clear Mode.
b Set the Start Mode (see page 34).
c Set the Stop Mode (see page 35).
d Set the Sample Interval (see page 37).
Setting the Clear Mode
To configure the Clear mode:
1
Select Logger | Sample Timing.
2 Under Clear Mode, select one of the following options:
Do not clear logger - This is the default setting if the
logger memory is not yet full. Use this setting when the
logger memory is not yet full and all you want to do is
change Stop mode settings without clearing the
existing samples. If the logger memory is full, or if Start
mode or Sample Interval changes are required, you