Configuring Loggers
Vaisala Veriteq vLog 4.5 43
5
For Timer Value, set the number of minutes or seconds
you want the transducer to begin warming up before
taking a sample. The default is 10 seconds.
6 Click OK.
Configuring Logger Channels to Work with
Thermocouples
This section applies only to those using Vaisala Veriteq
VL-1700 and SP-1700 Series Data Loggers. These loggers
work with various types of thermocouples (E, J, K, R, S, and
T). You need to configure your channel to specify the type of
thermocouple you are using.
To configure VL-1700 and SP-1700 Series Data Loggers to work
with thermocouples:
1
In vLog, choose Logger | Setup ().
2 From the Setup window, click Setup next to the channel
to configure.