Vaisala WINDCAP WMT52 Guía de usuario Pagina 65

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 143
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 64
Chapter 7 __________________________________________________ Getting the Data Messages
VAISALA_______________________________________________________________________ 63
Measurement Reset (aXZM)
This command is used to interrupt all ongoing measurements of the
sensor and start them from the beginning.
Command format in ASCII and NMEA 0183: aXZM<cr><lf>
Command format in SDI-12: aXZM!
Example (ASCII):
0XZM<cr><lf>
0TX,Measurement reset<cr><lf>
Example (SDI-12):
0XZM!0 (= device address)
Example (NMEA 0183):
0XZM<cr><lf>
$WITXT,01,01,09,Measurement reset*50<cr><lf>
where
a = Device address
XZM = Measurement break command
<cr><lf> = Command terminator in ASCII and NMEA 0183
! = Command terminator in SDI-12