Chapter 16 _____________________________________________________ Serial Line Commands
VAISALA ______________________________________________________________________ 107
HELP Outputting the Commands
Syntax: HELP<cr>
Example:
>help
Available commands:
ADDR CLOSE COOLER FLOW ICE INIE INTV
OPEN PRES PUMP R SALT SEND SMODE
SNUM START STOP TIME TP UNIT VERS
ECHO ALARM ASEL
SERI FORM ERRS ? DIR DEL PLAY
HELP
INIE Reverting the Factory Settings
Syntax: INIE<cr>
Does not affect calibration, date, time or data file memory.
Example:
>inie
All user settings will be deleted!
Are you sure : (Y/N) ? Y
OK
SMODE Selecting the Operation Mode
Syntax: SMODE x<cr>
where
x = STOP, RUN, or POLL
The SMODE command is used to set or inspect the default operation
mode of the EIA-232 (and optional RS485) interfaces.
STOP mode: After power-up the transmitter outputs its type and
software version and then waits for further commands.
RUN mode: Continuous output starts automatically from power-up.
Output can be defined with the command FORM. Only command S
can be used.
POLL mode: Allows the communication with multiple transmitters
or other digital instruments connected to one serial bus. Only the