Vaisala DEWCAP DM500 Guía de usuario Pagina 132

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 138
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 131
User's Guide ______________________________________________________________________
130 _________________________________________________________________ M210205EN-D
in air at that particular temperature. It is defined below as the ratio of
the water vapor pressure P
w
to the saturation water vapor pressure P
ws
.
%100/
=
wsw
PPRH (4)
Air temperature (ºC) Dewpoint temperature (ºC)
0102030405060708090100
100
80
60
40
20
0
-20
-40
1
6
0
140
120
100
80
60
40
20
0
-20
-40
Relative humidity (%RH)
Figure 26 Water Vapor Saturation Curves
Mixing Ratio (x)
Mixing ratio (x) is mass of water vapour per unit mass of dry air with
which it is associated.
wtot
w
PP
P
x
-
9907.621 = (5)
where
P
w
= water vapor pressure (Pa)
P
to
t
= total process pressure (Pa)
Vista de pagina 131

Comentarios a estos manuales

Sin comentarios