Vaisala DEWCAP DM500 Guía de usuario Pagina 122

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 138
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 121
User's Guide ______________________________________________________________________
120 _________________________________________________________________ M210205EN-D
-0.4
-0.3
-0.2
-0.1
0
0.1
0.2
0.3
0
.
4
-80 -70 -60 -50 -40 -30 -20 -10 0 10 20
Dew / Frostpoint [°C]
Deviation temperature [°C]
Specifications +
Specifications -
Calibration uncertainty (+2s)
Calibration uncertainty (-2s)
Repeatability (+2s)
Repeatability (-2s)
Figure 23 Dewpoint Temperature Measurement Repeatability
Response time (see the following graph)
Typical Response Time for 5 °C Abrupt
Dew / Frostpoint Change (Flow 1.0 slpm)
0
5
10
15
20
25
30
35
40
-70 -60 -50 -40 -30 -20 -10 0 10 20 30 40 50 60
Dew / Frostpoint [°C]
Time to ±0.5 °C from final reading
[minutes]
Figure 24 Dewpoint Temperature Measurement Response
Time
Vista de pagina 121
1 2 ... 117 118 119 120 121 122 123 124 125 126 127 ... 137 138

Comentarios a estos manuales

Sin comentarios