Vaisala DEWCAP DM500 Guía de usuario Pagina 124

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 138
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 123
User's Guide ______________________________________________________________________
122 _________________________________________________________________ M210205EN-D
Wetted parts' materials Stainless steel (AISI
316L), silicone
elastomer, Vectra LCP,
tantalum, quartz
Housing classification IP31 (NEMA 2)
Mechanical piping connections Swagelok 6 mm/ /¼"
Storage temperature range -40...+70 ºC
Weight 7 kg
Minimum heat removal capacity
of the cooling water system 250 W @ 5 ºC
Options for DMP501
Pump 0...1 l/min,
0... 200 mbar
differential pressure
Heated internal gas sampling
Connection cable to DMI500 2.5 or 10 meters
Temperature probe
Measurement range -40…+80 ºC
(-40…+176 ºF)
Typical accuracy ±0.1 ºC (±0.18 ºF)
Sensor Pt100 PRT DIN
IEC 751 class ¼ B
Cable length 2 m /5 m /10 m
DMI500 User Interface Unit
Operating voltage 100-230 VAC
Operating current 5A-3A
Operating power max 460 W
Display B/W LCD with
backlight
Menu languages English, German,
French, Finnish and
Spanish
Operating temperature 0...+40 ºC
Storage temperature range -40...+70 ºC
Electronics housing material PPE + PS plastic,
aluminium, stainless
steel
Housing classification IP31 (NEMA 2)
Weight 7.5 kg
Vista de pagina 123
1 2 ... 119 120 121 122 123 124 125 126 127 128 129 ... 137 138

Comentarios a estos manuales

Sin comentarios