Chapter 4 _________________________________________________________________ Operation
VAISALA_______________________________________________________________________ 103
Manual Sensor Purge
The Purge should be performed always before calibration (see the
calibration instructions) or when there is a reason to believe that a
sensor has become exposed to an interfering chemical.
1. Open the MAIN MENU by pressing any of the ▲▼◄► arrow
buttons.
2. Select Measuring, press the arrow button.
3. Select Purge, press the arrow button.
4. Select Manual Purge using the arrow buttons.
0504-050
Figure 48 Performing Manual Purge
5. Press START to start manual Purge. Press CANCEL to
interrupt the manual sensor Purge.
6. Press EXIT to return to the basic display.
Interval Purge
When DMT345 and DMT346 leave the factory the interval Purge
takes place repeatedly with the time intervals set in the factory. The
user can change the interval in which the purge takes place by using
serial commands or with the optional display/keypad. This can be
needed if the measuring environment contains high concentrations of
interfering chemicals.
Power-Up Purge
Purge is to start within 10 seconds from the power-up of the device.
This function can be disabled by using the serial line or the
display/keypad.