User's Guide _______________________________________________________________________
80 ___________________________________________________________________ M210762EN-A
Example:
>?
DMT340 / 3.01
Serial number : A3220001
Batch number : A3120029
Adjust. date : 2005-08-09
Adjust. info : Vaisala/HEL
Date : 2000-01-01
Time : 00:42:41
Serial mode : STOP
Baud P D S : 4800 E 7 1
Output interval: 0 s
Address : 0
Echo : ON
Pressure : 1000.00 hPa
Filter : ON
Ch1 output : 4...20mA
Ch2 output : 4...20mA
Ch1 Tdf low : 0.00 'C
Ch1 Tdf high : 100.00 'C
Ch2 x low : 0.00 g/kg
Ch2 x high : 1000.00 g/kg
Module 1 : not installed
Module 2 : not installed
>
HELP
Use the command HELP to list the commands.
>help
? ACAL ADDR AERR ASCL
ASEL CDATE CLOSE CODE CRH
CT CTA CTEXT DATE DELETE
DIR DSEL DSEND ECHO ERRS
FCRH FDATE FILT FORM FST
FTIME HELP INTV ITEST MODS
OPEN PLAY PRES R RESET
SEND SERI SMODE TEST TIME
UNDELETE UNIT VERS XPRES
>
ERRS
Use the command ERRS to display transmitter error messages, see
Table 28 on page 122.
Examples:
>ERRS
NO ERRORS
>