Chapter 7 _____________________________________________________________ Technical Data
VAISALA_______________________________________________________________________ 129
Table 33 Dewpoint Specifications DMT346
Property Description / Value
Sensor Vaisala DRYCAP®180S
Measurement range -25 … +100 °C (-13 … +212 °F) T
d
Accuracy ±2 °C (±3.6 °F) T
d
See the accuracy graph in Figure 55
below
Response time 63% [90%] flow rate
1l/min and 1 bar pressure
from dry to wet 5s [10 s]
from wet to dry including auto-
calibration 45s [5 min]
0604-022
Figure 55 Dewpoint Measurement Accuracy Graph for
Vaisala DRYCAP
®
Dewpoint Transmitter DMT346
Table 34 Mixing Ratio Specifications DMT346
Property Description / Value
Measurement range (typical) 0 ... 1000 g/kg (0 … 7000 gr/lbs)
Accuracy ±12 % of reading
Comentarios a estos manuales